Product Introduction:
Infant incubator is provided by TRIUP International Corp.,a major Medical Equipment supplier in Health & Beauty wholesale industry in China.You can check out the specification of Infant incubator below or get any optional result clicking into the following keywords incubator , Infant incubator , TRBI-3000P , Description , Features and you can check the authentications and testimonials of Infant incubator .
Description:
Optional Devices (At Buyers Selection):
P
Blue Light Phototherapy Unit, LED Digital Timer, Display therapy time & total time.
STANDARD COMPOSITION:
1Servo Control Mode: air temperature, skin temperature;
2Instant Parameter Displaying: set temp, air temp, skin temp, internal humidity, heat power rate;
3Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air & skin temp override, fan failure, system failure;
4>37 Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;
5Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf;
6RS-232 Output Connector (Use for network system).
Specifications:
1)Hood specifications:
2)Mattress Tray Size...36cm×65cm
3)Mattress to Hood..37cm
4)Access Panel Height.24cm
5)Standard Hood Includes: 4 port doors; 2 iris ports; front access panel; 2 left and 2 right tubing grommets.
6)Power supplyAC220V, 50Hz
7)Power requirement.≤450VA
8)Operational condition:
9)Environment temp+20+30
10)Environment relative humidity
....30%75%
11)Environment air velocity of flow<0.3m/s
12)Temperature fluctuation...≤0.5
13)Mattress temperature uniformity..≤0.8(level mattress)
14)Temperature rise time.≤30min
15)Noise level within hood environment..≤55dB (environment noise level≤45dB)
16)Mattress tray Max. load10kg
17)Transfusion pole Max. load....2kg
18)Shelf Max. load.10kg
19)Cabinet Max. load...6kg
20)Main machine:
Packing size..100 cm×58 cm×80cm;
Net weight.40kg; Gross Weight.55kg;
21)Machine stand:
Packing size..100 cm×58 cm×80cm;
Net weight...45kg; Gross Weight65kg;
Specifications:
Air Temp Control Range:
2037
Temperature override mode to:
39
Skin Temp Control Range:
3037
Temperature override mode to:
38
Skin temp sensor accuracy:
≤0.3
Humidity display range:
5%-99%
Alarms
Power failure alarm:
Yes
Fan failure alarm:
Yes
Sensor failure alarm:
Yes
High air temperature alarm:
+1.5
High skin temperature alarm:
+1.0
Low air temperature alarm:
-2.5
Low skin temperature alarm:
-1.0
Over air temperature alarm:
≤38.0; ≤40.0
Over skin temperature alarm:
≤38.0; ≤39.0
System failure alarm:
Yes
Stepless:
Yes